Answers are grounded in this paper and cite their sources.
Introduction In our study, we aimed to evaluate the analytical process performances of the biochemistry tests in the analysis systems that were widely used in the clinical laboratories by using the six-sigma methodology. Methods The analytical performances of four different analytical platforms (Beckman Coulter-Olympus AU2700, Abbott-Architect C8000, Roche-Cobas 8000, and Siemens-ADVIA 2400) running 18 biochemical tests (urea, creatinine, uric acid, total bilirubin, AST, ALT, ALP, LDH, HDL-C, Ca 2+ , cholesterol, TG, glucose, total protein, albumin, Na + , K + , and Cl − ) were evaluated. Biases and CV were obtained from internal quality control programs. The performance of each instrument was assessed using the σ scale; the benchmark performance was <3 σ (improvement needed); “world-class” quality was six sigma. Results The parameters that have σ≥6 which means in world class are HDL-C and ALP in all four systems, while only Na + has σ<3 which point out the need to develop. Similarities and differences were found between systems for analytical process sigma values of all tests except these. Discussion and conclusion To improvement and monitoring of the analytical process performance as a part of total quality of a clinical laboratory to provide continuous improving, sigma levels can be used as it is a reliable method.
No takes yet. Share an insight, caveat, or question.
Gülbahar et al. (2016) studied this question.