Answers are grounded in this paper and cite their sources.
Liquid-chromatography-tandem-mass-spectrometry (LC-MS/MS) is considered the gold standard for the measurement of vitamin D metabolites as it does not suffer from most of the limitations inherent to immunoassays. In addition, LC-MS/MS offers the advantage measuring several metabolites simultaneously. Despite many advantages, LC-MS/MS methods are not free from analytical challenges. Aspects such as sample type, protein precipitation, analyte extraction, derivatization, chromatographic separation ionization and capabilities of the mass spectrometer have to be considered carefully. Calibration, standardization and the use of internal standards are other important issues that affect the accuracy of results. Only well designed methods that are continuously controlled by internal and external quality control allow an accurate and stable measurement 25-hydroxy vitamin D (25(OH)D). Other metabolites, such as 1,25(OH)2D and 24,25(OH)2D, are not yet standardized and comparable results between laboratories are difficult to obtain. While an extensive number of original studies and review articles has addressed the performance of 25(OH)D immunoassays, the analytical challenges regarding the measurement of vitamin D metabolites by LC-MS/MS are not widely known. Therefore, the aim of this review is to provide a comprehensive overview about existing LC-MS/MS methods for the measurement of vitamin D metabolites and their analytical performance.
No takes yet. Share an insight, caveat, or question.
Zelzer et al. (2018) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: