The photopeak efficiency of a NaI (Tl) detector was determined experimentally for 23‐1333 keV photon energy by using polyester coated radioisotopes 22Na, 54Mn, 57Co, 60Co, 109Cd, 137Cs, and 133Ba. The energy resolution of the NaI (Tl) detector was also measured. The data were analyzed using the Genie‐2000 from Canberra. The experimental values were fitted to an analytical function of photon energy and an agreement was observed for the entire range of the studied energies. The results have shown that the NaI (Tl) detector is a particularly attractive detector for field application, as it works at room temperature and has a high probability of detecting gamma radiation.