Key result
Outpatient video capsule endoscopy achieves a ~95% completion rate safely in patients with cardiac devices.
Why the study?
Implantable cardiac devices are a relative contraindication to video capsule endoscopy due to concerns about interference, and gastrointestinal outcomes of VCE in these patients are minimally described.
Is video capsule endoscopy safe and effective in outpatients with implantable cardiac devices?
Observational (n=21)
No
Is video capsule endoscopy safe and effective in outpatients with implantable cardiac devices?
Video capsule endoscopy appears safe and effective in outpatients with implantable cardiac devices when using a telemetry monitoring protocol.
Supports VCE feasibility in outpatients with cardiac devices; hypothesis-generating and requires prospective validation before practice change.
BACKGROUND AND AIM: Implantable cardiac devices are a relative contraindication to video capsule endoscopy (VCE) because of concerns regarding interference. As a result of a lack of alternatives, some centers have adopted protocols to allow for VCE in these patients. There are minimal published descriptions of the gastrointestinal outcomes of these procedures. We investigated the completion rate and diagnostic yield of VCE carried out in outpatients with implantable cardiac devices. METHODS: We carried out a retrospective review of all VCE from April 2010 through March 2013 at our center. Patients that underwent VCE through a specialized protocol for outpatients with implantable cardiac devices were identified. The protocol used telemetry for cardiac monitoring during battery lifespan of the VCE. Demographic data, procedure indications, results and any procedural events were collected. RESULTS: Twenty-one patients met the study criteria, with 16 (76%) pacemakers, four (19%) implantable cardioverter defibrilla tors and one (5%) implantable hemodynamic monitor. Two (10%) devices were adjusted prior to VCE. Twenty (95%) were completed to the colon, with a median gastric transit time of 18 min and a small bowel transit time of 216 min. The procedure was diagnostic in 13 (62%) patients. One (5%) capsule retention event occurred. No cardiac events or video capture disturbances occurred. CONCLUSIONS: Video capsule endoscopy done on outpatients with implantable cardiac devices through a cardiac monitoring protocol resulted in a satisfactory completion rate and high diagnostic yield with no adverse cardiac events. A large prospective trial is needed to confirm these novel results.
No takes yet. Share an insight, caveat, or question.
Stanich et al. (2014) conducted an observational in Implantable cardiac devices (n=21). Video capsule endoscopy was evaluated on Completion rate. Video capsule endoscopy in outpatients with implantable cardiac devices achieved a 95% completion rate and 62% diagnostic yield with no adverse cardiac events.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: