Key result
Dialysis machines set to deliver a sodium concentration of 136 mmol/l actually delivered a higher concentration of 141.8±2.9 mmol/l by flame photometry and 142.5±2.4 mmol/l by ISE.
Observational (n=16)
No
Absolute Event Rate: 141.8% vs 136%
Dialysis machines set to deliver hyponatraemic dialysate actually deliver higher sodium concentrations than set, highlighting the need to measure delivered concentrations in clinical trials.
Authors
No takes yet. Share an insight, caveat, or question.
Dialysis sodium delivery exceeds programmed values; leaves open accuracy of hyponatremic dialysate effects in prior trials.
Ekbal et al. (2016) conducted an observational in Hemodialysis (n=16). Hemodialysis machines set to 136 mmol/l sodium vs. Set dialysate sodium concentration was evaluated on Delivered dialysate sodium concentration. Dialysis machines set to deliver a sodium concentration of 136 mmol/l actually delivered a higher concentration of 141.8±2.9 mmol/l by flame photometry and 142.5±2.4 mmol/l by ISE.