Answers are grounded in this paper and cite their sources.
A novel pulse compression technique is developed that improves the axial resolution of an ultrasonic imaging system and provides a boost in the echo signal-to-noise ratio (eSNR). The new technique, called the resolution enhancement compression (REC) technique, was validated with simulations and experimental measurements. Image quality was examined in terms of three metrics: the eSNR, the bandwidth, and the axial resolution through the modulation transfer function (MTF). Simulations were conducted with a weakly-focused, single-element ultrasound source with a center frequency of 2.25 MHz. Experimental measurements were carried out with a single-element transducer (f/3) with a center frequency of 2.25 MHz from a planar reflector and wire targets. In simulations, axial resolution of the ultrasonic imaging system was almost doubled using the REC technique (0.29 mm) versus conventional pulsing techniques (0.60 mm). The -3 dB pulse/echo bandwidth was more than doubled from 48% to 97%, and maximum range sidelobes were -40 dB. Experimental measurements revealed an improvement in axial resolution using the REC technique (0.31 mm) versus conventional pulsing (0.44 mm). The -3 dB pulse/echo bandwidth was doubled from 56% to 113%, and maximum range sidelobes were observed at -45 dB. In addition, a significant gain in eSNR (9 to 16.2 dB) was achieved.
No takes yet. Share an insight, caveat, or question.
Michael L. Oelze (2007) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: