Purpose: To evaluate a dual-implant system combining a capsular-bag monofocal intraocular lens with a sulcus-implanted small-incision new-generation implantable miniature telescope (SING-IMT). Methods: Optical simulations quantified retinal spot size, modulation transfer function (MTF), and tolerance to misalignment. Preclinical porcine experiments evaluated surgical feasibility, anatomical stability, and early safety. First-in-human (FiH) implantation assessed surgical outcomes, anterior segment behaviour using ultrasound biomicroscopy and optical coherence tomography, functional vision, and safety over 6 months. Results: Optical modelling demonstrated that the dual-implant configuration maintained retinal image quality within the rehabilitative regime for advanced AMD, with root-mean-square retinal spot radii of 18.7 µm (0°) and 20.4 µm (2.5°), below the estimated retinal sampling limit (∼49 µm), consistent with an expected visual acuity improvement of approximately 2 Snellen lines. Decentration up to 0.5 mm was optically well tolerated, whereas tilt produced rapid mid-frequency contrast loss. In vivo porcine studies confirmed stable sulcus seating without disruption of the capsular-bag IOL or anterior segment structures. FiH implantation was completed without intraoperative complications. At 6 months, distance visual acuity improved from 20/200 to 20/125, near acuity from 56 to 15 Colenbrander units, and reading speed from non-assessable preoperatively to 20 words/min, with no device-related adverse events. Conclusion: This integrated evaluation supports the feasibility and short-term safety of sulcus-based SING-IMT implantation as a minimally invasive intraocular magnification strategy for selected pseudophakic patients with advanced AMD.
No takes yet. Share an insight, caveat, or question.
Rizzo et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: