Colloid osmotic pressure (COP) of some of the most frequently used plasma replacement fluids was measured with a colloid osmometer. COP of 4% human albumin solutions was only half that of normal human serum (13.6 +/- 0.6 vs. 27.5 +/- 2.7 mmHg (1.8 +/- 0.1 vs. 3.7 +/- 0.4 kPa)) (mean +/- s.d.), whereas COP of 20% human albumin solutions was eight times higher (196.0 +/- 12.3 mmHg (26.1 +/- 1.6 kPa)). Enhancing the protein concentration from 4% to 20% in the human albumin solutions increased COP 14-fold, reflecting the exponential relationship between protein concentration and COP of a solution. Fresh donor plasma furnished by the hospital blood-bank had a COP about 30% below normal human serum (18.1 +/- 1.3 mmHg (2.4 +/- 0.2 kPa)), due to dilution during preparation. Dextran 70 (6%) had a COP more than twice, and Ringer-Dextran 60 (3%) about 75% of that of normal human serum. Dextran 40 (10%) and gelatin (3.5%, Haemaccel) leaked markedly through the membrane of the colloid osmometer, making acceptable measurements impossible. Seven different hydroxyethyl starch (HES) solutions were measured, and the COP varied between half and 3 times that of normal human serum, depending on molecular weight and concentration of the HES.
No takes yet. Share an insight, caveat, or question.
Tønnessen et al. (1993) studied this question.
Synapse has enriched one closely related paper. Consider it for comparative context: