Key result
LivaNova Inspire Min.I shows comparable air removal performance to Getinge in minimized ECC circuits.
Why the study?
Minimally invasive extracorporeal circulation systems lack a venous reservoir, potentially reducing air handling properties compared to conventional cardiopulmonary bypass systems.
Does the LivaNova Inspire Min.I system improve air removal compared to the Getinge minimized ECC system in an in vitro model?
Does the LivaNova Inspire Min.I system improve air removal compared to the Getinge minimized ECC system in an in vitro model?
Both the LivaNova Inspire Min.I and Getinge minimized ECC systems showed comparable air removal properties in an in vitro model, supporting their safety for clinical use.
Both MiECC systems show comparable in vitro air handling; leaves open whether marginal differences affect clinical embolization risk.
Minimally invasive extracorporeal circulation systems are developed to decrease the deleterious effects of cardiopulmonary bypass. For instance, prime volume and foreign surface area are decreased in these systems. However, because of the lack of a venous reservoir in minimized systems, air handling properties of these minimally invasive extracorporeal circulation systems may be decreased as compared to conventional cardiopulmonary bypass systems. The aim of this in vitro study is to compare the air handling properties of two complete minimized cardiopulmonary bypass systems of two manufacturers, of which one system is provided with the air purge control. In an in vitro study, two minimally invasive extracorporeal circulation systems, Inspire Min.I manufactured by Sorin Group Italia, Mirandola, Italy (LivaNova, London, United Kingdom) and minimized extracorporeal circulation manufactured by Maquet, Rastatt, Germany (Getinge, Germany), were challenged with two types of air challenges; a bolus air challenge and a gaseous microemboli challenge. The air removal characteristics of the venous bubble traps and of the complete minimally invasive extracorporeal circulation systems were assessed by measuring the gaseous microemboli volume and number downstream of the venous bubble traps in the arterial line with a bubble counter. No significant differences were observed in air reduction between the venous bubble traps of Getinge (venous bubble traps) and LivaNova (Inspire venous bubble traps 8 in conjunction with the air purge control). Similarly, no significant differences were observed in volume and number of gaseous microemboli in the arterial line of both complete minimally invasive extracorporeal circulation systems. However, the gaseous microemboli load of the Inspire Min.I system was marginally lower after both the bolus air and the gaseous microemboli challenges. Both minimally invasive extracorporeal circulation systems assessed in this study, the LivaNova Inspire Min.I and the Getinge minimized extracorporeal circulation, showed comparable air removal properties, after both bolus and gaseous microemboli air challenges. Besides, air purge control automatic air removal system provided with the LivaNova Inspire Min.I. system may enhance patient's safety with the use of a minimally invasive extracorporeal circulation system. We consider both systems equally safe for clinical use.
No takes yet. Share an insight, caveat, or question.
Stehouwer et al. (2019) studied this question. LivaNova Inspire Min.I system vs. Getinge minimized extracorporeal circulation system was evaluated on Air removal characteristics (gaseous microemboli volume and number downstream of venous bubble traps). In an in vitro comparison, the LivaNova Inspire Min.I and Getinge minimized extracorporeal circulation systems showed comparable air removal properties after bolus and gaseous microemboli challenges.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: