We report our evaluation of a commercially available procedure and reagents for determination of ferritin in serum by enzyme-labeled immunosorbent assay (ELISA). Results by our procedure and the "Fer Iron" (Ramco Labs) procedure shows a degree of association (r) of 0.95 and a regression equation of y = 1.03x - 33. Similarly, our procedure compared to the "Gamma Dab" (Clinical Assays) ferritin procedure shows a degree of association of 0.98 and a regression equation of y = 0.93x - 11. Between-day standard deviations were 6 and 22 micrograms/L (n = 24 and 20) for ferritin concentrations of 20 and 300 micrograms/L, respectively. Ferritin values showed no correlation with total iron concentration, but show a broad inverse relationship with iron-binding capacities. The favorable correlation with existing procedures and the speed of the analysis commend the use of ELISA for measurement of ferritin in serum.
No takes yet. Share an insight, caveat, or question.
Fortier et al. (1979) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: