Key result
Computerized autonomic testing correlates strongly with manual methods for measuring heart rate responses.
Why the study?
A standardized, objective, and faster computerized method for measuring heart rate and heart rate response to autonomic function tests was needed to replace the traditional manual method.
Does a computerized data analysis system provide comparable results to the traditional manual method for evaluating cardiovascular autonomic responses in healthy subjects and diabetic patients?
Observational (n=240)
Does a computerized data analysis system provide comparable results to the traditional manual method for evaluating cardiovascular autonomic responses in healthy subjects and diabetic patients?
A computerized method for calculating autonomic function test responses is comparable to the traditional manual method, offering advantages of speed and objectivity.
Answers are grounded in this paper and cite their sources.
Computerized analysis correlates strongly with manual HR/HRR methods; leaves open routine clinical adoption in diabetes pending larger validation studies.
The objectives of the present study were 1) to compare results obtained by the traditional manual method of measuring heart rate (HR) and heart rate response (HRR) to the Valsalva maneuver, standing and deep breathing, with those obtained using a computerized data analysis system attached to a standard electrocardiograph machine; 2) to standardize the responses of healthy subjects to cardiovascular tests, and 3) to evaluate the response to these tests in a group of patients with diabetes mellitus (DM). In all subjects (97 healthy and 143 with DM) we evaluated HRR to deep breathing. HRR to standing, HRR to the Valsalva maneuver, and blood pressure response (BPR) to standing up and to a sustained handgrip. Since there was a strong positive correlation between the results obtained with the computerized method and the traditional method, we conclude that the new method can replace the traditional manual method for evaluating cardiovascular responses with the advantages of speed and objectivity. HRR and BPR of men and women did not differ. A correlation between age and HRR was observed for standing (r = -0.48, P < 0.001) and deep breathing (r = -0.41; P < 0.002). Abnormal BPR to standing was usually observed only in diabetic patients with definite and severe degrees of autonomic neuropathy.
No takes yet. Share an insight, caveat, or question.
Neumann et al. (1997) conducted an observational in Diabetes mellitus and healthy subjects (n=240). Computerized data analysis system vs. Traditional manual method was evaluated on Heart rate and heart rate response to Valsalva maneuver, standing, and deep breathing. A computerized data analysis system strongly correlated with manual methods; age correlated with heart rate response to standing and deep breathing.