This is the authors' abstract. We don't add key points for this paper.
Interferometric wavelength meters have attained frequency resolutions down to the megahertz range. In particular, Fizeau interferometers, which have no moving parts, are becoming a popular tool for laser characterization and stabilization. In this paper, we characterize such a wavelength meter using an ultrastable laser in terms of relative frequency instability σ(y)(τ) and demonstrate that it can achieve a short-term instability σ(y)(1s)≈2×10(-10) and a frequency drift of order 10 MHz/day. We use this apparatus to demonstrate frequency control of a near-infrared laser, where a frequency instability below 3×10(-10) from 1 to 2000 s is achieved. Such performance is, for example, adequate for ion trapping and atom cooling experiments.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
A 2015 study studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: