Answers are grounded in this paper and cite their sources.
We report on the characterization of a closed-cycle top-loader at the Korea Research Institute of Standards and Science (KRISS) for quantum resistance metrology. The quantized Hall resistance (QHR) is realized with graphene below a magnetic field of 6 T at a base temperature of 4.5 K. The rotary valve of a pulse-tube cryocooler for the top-loader is remotely mounted in a massive acoustic enclosure to reduce vibrations. This suppresses vibration-induced current noise in a coaxial cable of a top-loading probe to smaller than 0.1 pA/√Hzat the fundamental frequency near 1 Hz and its harmonics below 10 Hz. The reliability of the graphene-based QHR in the top-loader is examined through a comparison with the GaAs-based KRISS QHR standard in a wet cryomagnet via a mobile 1-kΩresistance reference. The degree of equivalence is smaller than an expanded measurement uncertainty ($k =2$) of 7 nΩ/Ω.
No takes yet. Share an insight, caveat, or question.
Chae et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: