Key result
Intraoperative MRI surgeries carry ~16% safety incidents, predominantly staff-related with no major harm.
Why the study?
The study aimed to determine the types of safety incidents occurring in the iMRI operating theatre to prevent future avoidable incidents in this hazardous environment.
Observational (n=271)
No
A significant proportion of safety incidents in the iMRI operating theatre are preventable staff errors, highlighting the need for training and strict protocols.
Answers are grounded in this paper and cite their sources.
Preventable staff errors drove one-third of iMRI incidents without major harm; hypothesis-generating for targeted training protocols.
Introduction: The purpose of this study was to determine the types of incidents that occurred in the iMRI OT over a nineteen-month period in our institution. We aim to prevent any future avoidable incidents from happening in this potentially hazardous environment. Methods: This is a single centre prospective non-anonymous observational study conducted from February 2009 to September 2010 on surgeries performed in the iMRI OT. Safety incidents specific to the iMRI OT such as violation of safety protocols and equipment failures were reported as well as any other safety incidents resulting in potential or actual adverse safety outcomes. The outcomes of the incidents were included and the data analysed at the end of the study period. Results: Of 271 cases that were operated in the iMRI OT, 43 incidents were reported by the staff involved in the care of the patient. Of the 43 incidents, 14 incidents (32.6%) were classified as staff/personnel error and were preventable. Incidents resulted in either delayed surgery or cancellation of the surgery. There were no major adverse incidents that led to patient harm. Conclusion: Many of the incidents were preventable and measures have been instituted to prevent the recurrence of such incidents. Staff training, safety protocols and stringent maintenance of equipment are paramount to safe and efficient use of the iMRI operating theatre.
No takes yet. Share an insight, caveat, or question.
Tan et al. (2014) conducted an observational in Surgeries in the intraoperative MRI operating theatre (n=271). Intraoperative magnetic resonance imaging operating theatre (iMRI OT) was evaluated on Safety incidents specific to the iMRI OT. Among 271 surgeries performed in an intraoperative MRI operating theatre, 43 safety incidents were reported, of which 32.6% were preventable staff errors, with no major patient harm.
Synapse has enriched 10 related papers. Consider them for comparative context: