Answers are grounded in this paper and cite their sources.
We report the design and testing of a nuclear magnetic resonance (NMR) microcoil receiver apparatus, employing solenoidal microreceiver coils of dimensions of tens to hundreds of microns, using applied field of 9 T (proton resonance frequency 383 MHz). For the smallest receiver coils we attain sensitivity sufficient to observe proton NMR with signal to noise (S/N) one in a single scan applied to a ∼10 μm3 (10 fl) water sample, containing 7×1011 total proton spins. We also test the dependence of the S/N on important coil parameters, including coil composition and resistivity, turn spacing, and lead lengths.
No takes yet. Share an insight, caveat, or question.
Seeber et al. (2001) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: