Key result
Novel self-gated fetal cine MRI matches echocardiography diameter measurements within ~0.3 mm bias.
Why the study?
Established cine MRI methods are not suitable for fetal heart imaging in utero due to small anatomical structures and difficulty in ECG synchronization.
Does fetal cardiac cine MRI provide reliable cardiac measurements compared to echocardiography in utero?
Observational (n=6)
No
Does fetal cardiac cine MRI provide reliable cardiac measurements compared to echocardiography in utero?
A novel fetal cardiac cine MRI technique provides reliable in utero cardiac measurements comparable to echocardiography, offering potential for prenatal evaluation of congenital heart defects.
Answers are grounded in this paper and cite their sources.
Offers potential echocardiography alternative for fetal hearts; extends technical validation but should not yet change practice.
Fast magnetic resonance imaging (MRI) led to the emergence of 'cine MRI' techniques, which enable the visualization of the beating heart and the assessment of cardiac morphology and dynamics. However, established cine MRI methods are not suitable for fetal heart imaging in utero, where anatomical structures are considerably smaller and recording an electrocardiogram signal for synchronizing MRI data acquisition is difficult. Here we present a framework to overcome these challenges. We use methods for image acquisition and reconstruction that robustly produce images with sufficient spatial and temporal resolution to detect the heart contractions of the fetus, enabling a retrospective gating of the images and thus the generation of images of the beating heart. To underline the potential of our approach, we acquired in utero images in six pregnant patients and compared these with their echocardiograms. We found good agreement in terms of diameter and area measurements, and low inter- and intra- observer variability. These results establish MRI as a reliable modality for fetal cardiac imaging, with a substantial potential for prenatal evaluation of congenital heart defects.
No takes yet. Share an insight, caveat, or question.
Chaptinel et al. (2017) conducted an observational in Fetal cardiac imaging (n=6). Fetal cardiac cine MRI with data-driven self-gating vs. Echocardiography was evaluated on Agreement of diameter measurements between MRI and echocardiography (Observer 1 bias) (Bias 0.30 mm, 95% CI -1.87; 2.47, p=0.266). Fetal cardiac cine MRI using a novel self-gating approach produced high-quality images of the beating heart with good agreement to echocardiography for diameter measurements (bias 0.30 mm).