Key result
Novel R-R interval algorithm achieves ~99% F1 score for wearable ECG detection during exercise.
Why the study?
Wearable ECG devices often contain noise and artifacts that hamper accurate extraction of QRS complexes, necessitating an R-R interval estimation algorithm tolerant to measurement faults for HRV analysis.
A novel R-R interval estimation algorithm using complex wavelet filtering improves QRS detection accuracy in noisy wearable ECG recordings.
Supports noise-tolerant RRI algorithms for wearable ECGs; leaves open prospective clinical validation before adoption.
Recent innovations in wearable electrocardiogram (ECG) devices have enabled various personal healthcare applications based on heart rate variability (HRV). However, wearable ECGs rarely undergo visual inspection by medical experts, hence may contain noise and artifacts. Because apparent changes in the recorded ECGs caused by noise and artifacts may hamper the extraction of QRS complexes, an R-R interval (RRI) estimation algorithm tolerant to these measurement faults is required as the initial step toward HRV analysis using wearable ECGs. This paper proposes a semi-real-time RRI estimation for wearable ECGs utilizing a two-stage structure. In the preprocessing stage, we use a complex-valued wavelet that can adaptively fit to morphological variations of the QRS complex while retaining computing resources for extracting the QRS complex features. In the decision stage, we make use of complex-valued features and select appropriate QRS complexes in consideration of three features: peak magnitude, peak location, and peak morphology (phase). Initial evaluations show that the QRS complex detection performance of the proposed method achieved the F1 score of 0.952 ± 0.040 when targeting pseudo ECG data created from open data assuming wearable ECGs, and of 0.986 ± 0.018 when targeting actual ECG data recorded by a shirt-type wearable ECG device during an exercise activity. Furthermore, the proposed method was able to suppress overlook or misdetection of QRS complexes, so the obtained RRIs are closer to the reference RRIs. The proposed method therefore contributes to achieving accurate HRV analysis using wearable ECGs in terms of obtaining accurate RRIs.
No takes yet. Share an insight, caveat, or question.
Shimauchi et al. (2021) studied ECG signal processing (n=3). Proposed R-R interval estimation algorithm vs. Pan-Tompkins algorithm was evaluated on F1 score for QRS complex detection. The proposed R-R interval estimation algorithm achieved an F1 score of 0.952 on pseudo ECG data and 0.986 on actual wearable ECG data during exercise, improving robustness against noise and artifacts.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: