Answers are grounded in this paper and cite their sources.
Endotoxins are lipopolysaccharide components of the cell membrane of Gram-negative bacteria that trigger the body's innate immune system and can cause shock and death. Water for medical therapy, including parenteral and dialysate solutions, must be free of endotoxin. This purity is challenging to achieve as many Gram-negative bacteria are endemic in the environment, and can thrive in harsh, nutrient-poor conditions. Current methods for removing endotoxin include distillation and reverse osmosis, both of which are resource intensive processes. Membranes that present an absolute barrier to macromolecular passage may be capable of delivering pure water for biomedical applications. In this work, endotoxin has been filtered from aqueous solutions using silicon nanopore membranes (SNMs) with monodisperse pore size distributions. SNMs with critical pore sizes between 26 and 49 nm were challenged with solutions of deionized water spiked with endotoxin and with Pseudomonas cepacia. The filtrate produced by the SNM from Pseudomonas-contaminated water had <1.0 endotoxin unit (EU) ml−1, which meets standards for dialysate purity. This approach suggests a technique for single-step cleanup of heavily contaminated water that may be suitable for field or clinical use.
No takes yet. Share an insight, caveat, or question.
Smith et al. (2011) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: