Indicator-dilution reliably measures regurgitant flow regardless of residual volume or mixing degree.
Why the study?
The conditions required for valid measurement of regurgitant flow across incompetent valves by indicator-dilution were not established.
This study establishes the theoretical and practical conditions required for valid measurement of valvular regurgitation using indicator-dilution techniques.
Answers are grounded in this paper and cite their sources.
Confirms indicator-dilution validity under specified conditions in animal models; leaves open translation to clinical regurgitation assessment.
A theory is presented for the measurement of regurgitant flow across an incompetent valve by injection of indicator distal to the valve and recording of time concentration curves proximal and distal to the valve. The conditions which must be met for valid measurements are examined and demonstrated by a functioning mechanical model of the heart. The method is valid regardless of the size of the residual volumes or degree of mixing with these volumes if a representative sample of the concentration of indicator in the blood moving across the valve in both directions is obtained.
No takes yet. Share an insight, caveat, or question.
LACY et al. (1959) studied Incompetent valves. Indicator-dilution measurement was evaluated on Validity of regurgitant flow measurement. Measurement of regurgitant flow by indicator-dilution is valid regardless of residual volume size or mixing degree, provided a representative sample of indicator concentration is obtained.