Answers are grounded in this paper and cite their sources.
Loop-mediated isothermal amplification (LAMP) amplifies a target gene with high specificity and rapidity under isothermal conditions. LAMP assays were developed for the rapid detection of metallo-beta-lactamase (MBL) genes such as bla(IMP-1)) and bla(VIM-2). We initially designed specific primers to detect MBL genes for LAMP assays and evaluated the specificity and sensitivity of these assays. LAMP assays amplified MBL genes under a constant temperature of 63 degrees C within 1 hour, and were compared to PCR in MBL-producing strains. The results of MBL genes typing by LAMP assays agree completely with PCR results. The lower detection limits of bla(IMP-1)- and bla(VIM-2)-LAMP assays using real-time turbidimeters were 30cfu/test and 3cfu/test. After amplification, products were directly observed by the naked eye with a fluorescent detection reagent. In conclusion, LAMP assays are convenient, rapid, and fully feasible for detecting MBL genes in ordinary clinical microbiology laboratories without special apparatus.
No takes yet. Share an insight, caveat, or question.
Kojima et al. (2006) studied this question.