Key result
Doppler opening pressure correlates strongly with invasive mean arterial pressure in CF-VAD patients.
Why the study?
Accurate blood pressure measurement in continuous-flow ventricular assist device patients is imperative to reduce stroke risk, but the accuracy of the Doppler opening pressure method compared with arterial line measurement was unclear.
Does Doppler opening pressure measurement accurately estimate mean arterial pressure compared to arterial line measurement in CF-VAD patients?
Cohort (n=154)
Does Doppler opening pressure measurement accurately estimate mean arterial pressure compared to arterial line measurement in CF-VAD patients?
Doppler opening pressure accurately estimates mean arterial pressure in CF-VAD patients and is recommended as the standard non-invasive method for BP measurement in this population.
Supports Doppler opening pressure as reliable noninvasive MAP surrogate in CF-VAD patients; leaves open impact on stroke outcomes.
AIMS: Accurate blood pressure (BP) measurement in continuous-flow ventricular assist device (CF-VAD) patients is imperative to reduce stroke risk. This study assesses the accuracy of the Doppler opening pressure method compared with the gold standard arterial line method in CF-VAD patients. METHODS AND RESULTS: In a longitudinal cohort of HeartMate II and HVAD patients, arterial line BP and simultaneously measured Doppler opening pressure were obtained. Overall correlation, agreement between Doppler opening pressure and arterial line mean vs. systolic pressure, and the effect of arterial pulsatility on the accuracy of Doppler opening pressure were analysed. A total of 1933 pairs of Doppler opening pressure and arterial line pressure readings within 1 min of each other were identified in 154 patients (20% women, mean age 55 ± 15, 50% HeartMate II and 50% HVAD). Doppler opening pressure had good correlation with invasive mean arterial pressure (r = 0.742, P < 0.0001) and more closely approximated mean than systolic BP (mean error 2.4 vs. -8.4 mmHg). Arterial pulsatility did not have a clinically significant effect on the accuracy of the Doppler opening pressure method. CONCLUSIONS: Doppler opening pressure should be the standard non-invasive method of BP measurement in CF-VAD patients.
No takes yet. Share an insight, caveat, or question.
Li et al. (2019) conducted a cohort in Continuous-flow ventricular assist device (CF-VAD) (n=154). Doppler opening pressure method vs. Arterial line method was evaluated on Correlation of Doppler opening pressure with invasive mean arterial pressure (r = 0.742, p=<0.0001). Doppler opening pressure demonstrated good correlation with invasive mean arterial pressure (r = 0.742, P < 0.0001) and closely approximated mean blood pressure in CF-VAD patients.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: