Key result
FloTrac version 3.02 strongly correlates with intermittent cardiac output and accurately predicts fluid responsiveness.
Why the study?
The FloTrac system provides continuous arterial pressure-based cardiac output and stroke volume variation monitoring without calibration, but its role in optimizing fluid management alongside other hemodynamic parameters was unclear.
Observational (n=80)
The FloTrac system provides continuous cardiac output and stroke volume variation monitoring without calibration, which is useful for optimizing fluid therapy and oxygen balance when combined with CVP/PAOP and ScvO2/SvO2.
Answers are grounded in this paper and cite their sources.
Hypothesis-generating for FloTrac v3.02-guided fluid management; leaves open routine adoption without randomized confirmation.
フロートラックシステムはFloTracTM sensorとVigileoTM monitorからなるキャリブレーションを必要としない連続的動脈圧心拍出量モニター(arterial pressure-based cardiac output:APCO)である.この精度はVersion 3.02では間欠的熱希釈式心拍出量(intermittent cardiac output:ICO)測定値と高い相関を示した.またこのモニターは付加価値として1回拍出量変動(stroke volume variation:SVV)が測定できるが,このSVVは輸液反応性の指標としては有用であると考えられた.しかしAPCOとSVVでは輸液過剰の指標とはならないため,APCOとSVVで循環血液量を適正化し,その後中心静脈圧(central venous pressure:CVP)や肺動脈閉塞圧(pulmonary artery occlusion pressure:PAOP)をモニタリングしながら輸液量をコントロールする方法が望ましいと考えられた.またフロートラックシステムは中心静脈血酸素飽和度(ScvO2)もしくは混合静脈血酸素飽和度(SvO2)が測定できる.これらを前述したAPCO,SVVやCVP,PAOPなどと併せて用いることにより最適な酸素需給バランスを保つことが可能であると考えられた.
No takes yet. Share an insight, caveat, or question.
Isao Fukuda (2011) conducted an observational in Surgical patients requiring cardiac output monitoring (n=80). FloTrac system (APCO and SVV) vs. Pulmonary artery catheter (ICO, CVP, PAOP) was evaluated on Correlation between arterial pressure-based cardiac output (APCO Ver. 3.02) and intermittent cardiac output (ICO) (r=0.75). The FloTrac system software version 3.02 demonstrated a strong correlation with intermittent cardiac output (r=0.75), and stroke volume variation accurately predicted fluid responsiveness (ROC area 0.93).