Key result
NeoBeat dry-electrode ECG matches conventional 3-lead monitoring for newborn heart rate detection.
Why the study?
The accuracy of heart rate detection by a novel wireless dry-electrode ECG device compared to conventional 3-lead gel-electrode ECG in newborns was not established.
Does a wireless, dry-electrode ECG device accurately detect heart rate compared to a conventional 3-lead gel-electrode ECG monitor in newborns?
Observational (n=50)
No
Does a wireless, dry-electrode ECG device accurately detect heart rate compared to a conventional 3-lead gel-electrode ECG monitor in newborns?
A novel wireless dry-electrode ECG device provides accurate heart rate detection comparable to conventional gel-electrode ECG in newborns.
Answers are grounded in this paper and cite their sources.
May support wireless newborn monitoring in research; leaves open clinical adoption pending validation in broader populations.
Objective To compare the accuracy of heart rate detection properties of a novel, wireless, dry-electrode electrocardiogram (ECG) device, NeoBeat®, to that of a conventional 3-lead gel-electrode ECG monitor (PropaqM®) in newborns. Results The study population had a mean gestational age of 39 weeks and 2 days (1.5 weeks) and birth weight 3528 g (668 g). There were 950 heart rate notations from each device, but heart rate was absent from the reference monitor in 14 of these data points, leaving 936 data pairs to compare. The mean (SD) difference when comparing NeoBeat to the reference monitor was -0.25 (9.91) beats per minute (bpm) (p = 0.44). There was a deviation of more than 10 bpm in 7.4% of the data pairs, which primarily (78%) was attributed to ECG signal disturbance, and secondly (22%) due to algorithm differences between the devices. Excluding these outliers, the correlation was equally consistent (r2 = 0.96) in the full range of heart rate captured measurements with a mean difference of − 0.16 (3.09) bpm. The mean difference was less than 1 bpm regardless of whether outliers were included or not.
No takes yet. Share an insight, caveat, or question.
Pike et al. (2021) conducted an observational in Healthy newborns (n=50). NeoBeat dry-electrode ECG vs. PropaqM conventional 3-lead gel-electrode ECG was evaluated on Mean difference in heart rate detection (MD -0.25 bpm, p=0.44). The NeoBeat dry-electrode ECG accurately detected heart rate in healthy newborns with a mean difference of -0.25 beats per minute compared to a conventional 3-lead gel-electrode ECG.