A full aperture backscatter station (FABS) target diagnostic has been activated on the first four beams of the National Ignition Facility. FABS measures both stimulated Brillouin scattering and stimulated Raman scattering with a suite of measurement instruments. Digital cameras and spectrometers record spectrally resolved energy for both P and S polarized light. Streaked spectrometers measure the spectral and temporal behavior of the backscattered light. Calorimeters and fast photodetectors measure the integrated energy and temporal behavior of the light, respectively. This article provides an overview of the FABS measurement system and detailed descriptions of the diagnostic instruments and the optical path.