Integrating sphere light sources that can deliver spectral radiance over several orders of magnitude are commonly employed to realize low-light-level (LLL) spectral radiance scales. In order to achieve accurate calibration of LLL measurement, a spectroradiometer with nonlinearity correction is used to measure the relative spectral distributions of a commercial integrating sphere light source. When the aperture of the light source is adjusted such that the integrated spectral radiance over 400–1000 nm changes from 3.91×10 2 W/(m 2 ⋅sr) to 3.14×10 −5 W/(m 2 ⋅sr), the relative spectral change exceeds 400% for the wavelength above 750 nm, and the corresponding luminance changes from 2.53×10 4 cd/m 2 to 2.97×10 −4 cd/m 2 . Multiple reflections between the adjustable aperture and the integrating sphere, and edge-scattered light at the aperture may introduce modulation effects on the relative spectral distribution. By introducing a new, to the best of our knowledge, integrating sphere light source structure, the relative spectral variation for nearly six orders of magnitude is kept within ±4.0%, which can greatly reduce the measurement uncertainty for LLL spectral radiance.