Key result
Mission Cholesterol point-of-care system shows high agreement with lab lipid testing, including ~0.97 triglyceride ICC.
Why the study?
Comprehensive analytical validation of point-of-care lipid testing devices against reference laboratory methods remains limited.
Does the Mission® Cholesterol 3-in-1 system provide accurate lipid profile measurements compared to reference laboratory methods in participants?
Population
120 participants
Comparison
Mission Cholesterol 3-in-1 system vs reference laboratory method
Design
Prospective cross-sectional agreement study
Authors
Loading...
The Mission® Cholesterol 3-in-1 point-of-care system demonstrates good analytical agreement and excellent diagnostic performance compared to reference laboratory methods, supporting its use as a rapid screening tool.
Cross-Sectional (n=120)
Does the Mission® Cholesterol 3-in-1 system provide accurate lipid profile measurements compared to reference laboratory methods in participants?
Effect estimate: ICC 0.972 for triglycerides
The Mission® Cholesterol 3-in-1 point-of-care system demonstrates good analytical agreement and excellent diagnostic performance compared to reference laboratory methods, supporting its use as a rapid screening tool.
Guzmán-Aquino et al. (2026) reported a cross-sectional. Mission® Cholesterol 3-in-1 system vs. reference laboratory method was evaluated on Analytical agreement of lipid profile measurements (ICC 0.972 for triglycerides). The Mission® Cholesterol 3-in-1 system showed good analytical agreement with laboratory testing, including an ICC of 0.972 for triglycerides, supporting its use as a rapid point-of-care tool.