Key result
Stylet-driven leads yield similar implant success rates to lumen-less leads for LBBAP.
Why the study?
Left bundle branch area pacing (LBBAP) has been performed exclusively using lumen-less pacing leads, and the safety and feasibility of LBBAP using stylet-driven leads had not been explored in a multicenter population.
Does left bundle branch area pacing using stylet-driven leads provide comparable safety and feasibility to lumen-less leads in patients requiring anti-bradycardia pacing or heart failure therapy?
Cohort (n=412)
Yes
Does left bundle branch area pacing using stylet-driven leads provide comparable safety and feasibility to lumen-less leads in patients requiring anti-bradycardia pacing or heart failure therapy?
Absolute Event Rate: 94% vs 88%
p-value: p=0.058
Left bundle branch area pacing using stylet-driven leads is safe and feasible, achieving high implant success rates and stable pacing thresholds comparable to traditional lumen-less leads.
Supports SDL feasibility for LBBAP; leaves open randomized trials on long-term outcomes versus LLL.
Funding Acknowledgements Type of funding sources: None. Background Left bundle branch area pacing (LBBAP) has been performed exclusively using lumen-less pacing leads (LLL) with fixed helix design. This study is the first to explore in a multicenter population the safety and feasibility of LBBAP using stylet-driven leads (SDL) with extendable helix design. Methods This study prospectively enrolled all consecutive patients who underwent LBBAP for anti-bradycardia pacing or heart failure indications at 8 Belgian hospitals. LBBAP was attempted either using SDL or LLL delivered through dedicated delivery sheaths. Implant success, complications, procedural and pacing characteristics were recorded at implant and follow-up. Results The study enrolled 412 patients (mean age 74 ± 38 years, 44% female). LBBAP with SDL and LLL was successful in 334/353 (94%) and 52/59 (88%) respectively (p=0.058). Implant success for LBBAP using SDL varied from 93 to 100% among centers. Mean paced QRS duration and stimulus to left ventricular activation time were comparable for SDL and LLL (126 ± 21ms versus 125 ± 22ms, p=0.861 and 74 ± 17 and 75 ± 21ms, p=0.756). SDL LBBAP resulted in low pacing thresholds (0.6 ± 0.4V at 0.4ms), which remained stable at 1, 6 and 12 months of follow-up. Lead revisions for SDL LBBAP occurred in 5(1.4%) patients: 2 dislodgements at day 1, 1 late septal perforation at week 3, 1 Twiddler’s syndrome at 5 months, 1 lead fracture at 6 months. Conclusion This multicenter study confirms the safety and feasibility of LBBAP using SDL, which is characterized by high implant success, low complication rate and low and stable pacing thresholds.
No takes yet. Share an insight, caveat, or question.
Pooter et al. (2022) conducted a cohort in Anti-bradycardia pacing or heart failure indications (n=412). Left bundle branch area pacing using stylet-driven leads vs. Lumen-less pacing leads was evaluated on Implant success (p=0.058). Left bundle branch area pacing using stylet-driven leads achieved a 94% implant success rate compared to 88% with lumen-less leads (p=0.058), with low and stable pacing thresholds.