Key result
Venus P-Valve successfully treats dysfunctional balloon-expandable pulmonary valve in valve-in-valve implantation.
Why the study?
Valve-in-valve transcatheter pulmonary valve replacement is established for degenerated pulmonary valves, but use of self-expanding valves within balloon-expandable valves is extremely limited.
Does valve-in-valve implantation of a self-expanding Venus P-Valve within a dysfunctional balloon-expandable pulmonary valve result in a competent valve in a patient with repaired tetralogy of Fallot?
Case Report (n=1)
Does valve-in-valve implantation of a self-expanding Venus P-Valve within a dysfunctional balloon-expandable pulmonary valve result in a competent valve in a patient with repaired tetralogy of Fallot?
This first-in-human case demonstrates that a self-expanding Venus P-Valve can be successfully implanted within a dysfunctional balloon-expandable pulmonary valve in selected patients.
May expand valve-in-valve options for degenerated balloon-expandable pulmonary valves; leaves open durability and safety pending prospective data.
BACKGROUND: Patients with repaired tetralogy of Fallot often require pulmonary valve interventions. Valve-in-valve transcatheter pulmonary valve replacement is an established strategy for degenerated pulmonary valves, but self-expanding valves within previously implanted balloon-expandable valves remain extremely limited. FIRST-IN-HUMAN/EARLY REPORTS SUMMARY: A 26-year-old man with repaired tetralogy of Fallot and prior Myval transcatheter heart valve (Meril Life Sciences Pvt. Ltd) implantation presented with exercise intolerance, severe pulmonary regurgitation, and right ventricular dilation. After computed tomography-based planning, redilation of the prosthetic valve, a Venus P-Valve was implanted within the dysfunctional balloon-expandable valve, resulting in a competent valve. DISCUSSION: This case suggests that, in selected regurgitant failures with adequate landing-zone length and successful scaffold preparation, a balloon-expandable valve may serve as a suitable platform for self-expanding valve-in-valve implantation. NOVELTY: This is, to our knowledge, the first Venus P-Valve implanted within a previously implanted balloon-expandable pulmonary valve. TAKE-HOME MESSAGE: Venus P-Valve implantation may be used as a transcatheter valve-in-valve solution in selected cases beyond native right ventricular outflow tract anatomies.
No takes yet. Share an insight, caveat, or question.
Güzeltaş et al. (2026) conducted a case report in Repaired tetralogy of Fallot with dysfunctional balloon-expandable pulmonary valve (n=1). Venus P-Valve implantation was evaluated on Valve competence. Implantation of a self-expanding Venus P-Valve within a dysfunctional balloon-expandable pulmonary valve was successfully performed, resulting in a competent valve.