Why the study?
Precise localization of non-pulmonary-vein ectopy using a single voltage-map dataset without rhythm-specific settings was needed to improve mapping accuracy.
Case Report (n=4)
Answers are grounded in this paper and cite their sources.
A stepwise mapping workflow using cycle-length-based separation and voltage-per-second filtering on the Affera system effectively localized non-pulmonary-vein ectopy with a median accuracy of 7 mm.
On the Affera system, a single voltage-map dataset containing both sinus and ectopic beats, collected without rhythm-specific settings, can be separated by cycle length (coupling interval) into SR and PAC activation maps. Stepwise filtering by activation threshold (voltage-per-second, excluding far-field potentials) and close-unipolar QS-like morphology then precisely localizes the ectopic origin.
No takes yet. Share an insight, caveat, or question.
Nishiuchi et al. (2026) conducted a case report in Paroxysmal atrial fibrillation with non-pulmonary-vein ectopy (n=4). Stepwise localization using cycle-length-based activation map separation and high V/s filtering was evaluated on Distance from identified mapping site to successful ablation site (median 7 mm).
Synapse has enriched 10 related papers. Consider them for comparative context: