Chemiresistive gas sensors are broadly cross-sensitive, limiting quantitative use in complex mixtures. We address this challenge through materials-and-array co-design: a 480-node library of composition-tuned, dendritic Pt-based nanoparticles (Pt, Pt-Ni, Pt-Ag, Pt-Pd, Pt-Cu, and Pt-Au) was challenged with twelve medically and environmentally relevant gases at 25 °C-namely isoprene, methanol, ethanol, acetone, ammonia, formaldehyde, NO, CO, CO2, H2, CH4, and C3H8-across 1 ppb to 100 ppm. Distinct composition-dependent response fingerprints emerged: Pt-Ni produced the largest formaldehyde response observed here (approximately 2.1% at 100 ppm), whereas Pt-Pd showed the strongest preference for NO (approximately 2.9% at 50 ppm). The films formed homogeneous, percolated networks that delivered reproducible signals, with device-to-device RSD < 4.2% across 480 sensors and >90% response retention over one month at room temperature. These results establish a practical route to heater-free gas sensing through bimetallic composition and array-level response profiling, while identifying humidity validation and multivariate classification as important next steps toward practical breath and environmental applications.
No takes yet. Share an insight, caveat, or question.
Barzegar et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: