Dissolved elemental oxygen is determined in a flowing aqueous stream using glucose oxidase to catalyse the reaction between D-glucose and O2 to produce hydrogen peroxide. The levels of the resulting H2O2 are detected and quantified by luminol chemiluminescence using in-line solid phase media for pH adjustment of the reagent stream and for controlled release of the luminophore. The reaction is initiated by electrochemical catalysis. By the use of excess D-glucose in the reagent flow stream, the intensity of chemiluminescence is rendered proportional only to fluctuations in the dissolved O2 concentration. The methodology provides a means for the detection of aqueous O2 in the range 0-10 mg/L.
No takes yet. Share an insight, caveat, or question.
Atwater et al. (1998) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: