Implementation of the AAPM Task Group 43 dosimetric formalism for 192Ir wires used as interstitial sources in low dose-rate (LDR) brachytherapy applications is investigated. Geometry factors, dose-rate constant values, radial dose functions, and anisotropy functions, utilized in this formalism, were calculated for various lengths of all commercially available wire source designs by means of a well-established Monte Carlo simulation code and an improved modification of the Sievert integral method. Results are presented in the form of look up tables that allow interpolation for dose-rate calculations around all practically used wire lengths, with accuracy acceptable for clinical applications.
No takes yet. Share an insight, caveat, or question.
Karaiskos et al. (2001) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: