Key result
Artificial neural network filtering reduces noise with less ECG distortion than conventional techniques.
Why the study?
ECG recordings often contain various noise interferences that limit their diagnostic utility and require effective removal methods.
The electrocardiogram (ECG) is the most widely used method for diagnosis of heart diseases, where a good quality of recordings allows the proper interpretation and identification of physiological and pathological phenomena. However, ECG recordings often have interference from noises including thermal, muscle, baseline and powerline noises. These signals severely limit ECG recording utility and, hence, have to be removed. To deal with this problem, the present paper proposes an artificial neural network (ANN) as a filter to remove all kinds of noise in just one step. The method is based on a growing ANN which optimizes both the number of nodes in the hidden layer and the coefficient matrices, which are optimized by means of the Widrow-Hoff delta algorithm. The ANN has been trained with a database comprising all kinds of noise, both from synthesized and real ECG recordings, in order to handle any noise signal present in the ECG. The proposed system improves results yielded by conventional techniques of ECG filtering, such as FIR-based systems, adaptive filtering and wavelet filtering. Therefore, the algorithm could serve as an effective framework to substantially reduce noise in ECG recordings. In addition, the resulting ECG signal distortion is notably more reduced in comparison with conventional methodologies. In summary, the current contribution introduces a new method which is able to suppress all ECG interference signals in only one step with low ECG distortion and a high noise reduction.
No takes yet. Share an insight, caveat, or question.
Mateo et al. (2012) studied Heart diseases (ECG recordings). Artificial neural network (ANN) filter vs. Conventional techniques of ECG filtering (FIR-based systems, adaptive filtering, wavelet filtering) was evaluated on Noise reduction and ECG signal distortion. An artificial neural network filter suppressed all ECG interference signals in a single step, yielding lower ECG distortion and higher noise reduction compared to conventional filtering techniques.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: