Key result
Novel aimed bipolar leads selectively isolate regional cardiac signals in a torso model.
Why the study?
Current ECG and VCG leads record combined signals from all cardiac regions, potentially cancelling important localized information that aimed leads could reveal.
Answers are grounded in this paper and cite their sources.
All currently employed ECG and VCG leads vectorially record added contributions from all cardiac regions. Aimed leads, if feasible, would yield information cancelled during addition. Three artificial dipoles (µ 1 , µ 2 , µ 3 ) are placed within a torso model in the approximate right ventricular free wall, posterobasal wall of left ventricle, and inferior cardiac surface, respectively. Dipoles are excited in turn and torso surface potential fields are mapped for each dipole. A bipolar lead connecting any pair of points on the same equipotential of the field produced by µ 1 does not show a deflection when µ 1 is excited, but responds to µ 2 and µ 3 . If this lead is now so placed that its two electrodes, while remaining on the µ 1 equipotential, are on the same equipotential of the µ 2 field as well, then it will respond to µ 3 alone; it is aimed at the latter. Superimposing the three field maps and searching for those equipotential lines of one map which cross the same equipotential line of another map twice, provides a large number of suitable leads, each aimed at one of the three dipoles. Five leads are described and tested in detail.
No takes yet. Share an insight, caveat, or question.
Fischmann et al. (1966) studied this question. Aimed bipolar leads was evaluated on Development and testing of aimed leads. Using a torso model with artificial dipoles, five aimed bipolar leads were developed and tested to selectively record contributions from specific cardiac regions.
Synapse has enriched 10 related papers. Consider them for comparative context: