Key result
Ungated radial dynamic perfusion CMR detects CAD with ~92% sensitivity and specificity.
Why the study?
Myocardial perfusion CMR is well-established for detecting ischemic heart disease, but ECG gating problems can cause image degradation and non-diagnostic scans, particularly in patients with arrhythmias.
Does rapid ungated or retrospectively self-gated radial myocardial perfusion CMR accurately detect coronary artery disease and provide high image quality in patients undergoing coronary angiography, including those with atrial fibrillation?
Cohort (n=8)
Blinded reader evaluation
No
Does rapid ungated or retrospectively self-gated radial myocardial perfusion CMR accurately detect coronary artery disease and provide high image quality in patients undergoing coronary angiography, including those with atrial fibrillation?
Answers are grounded in this paper and cite their sources.
Rapid ungated radial perfusion CMR provides high image quality and preliminary diagnostic accuracy for coronary artery disease even during atrial fibrillation, potentially expanding noninvasive stress CMR to patients with gating difficulties or arrhythmias.
BACKGROUND: Myocardial perfusion cardiovascular magnetic resonance (CMR) is a well-established method for detection of ischemic heart disease. However, ECG gating problems can result in image degradation and non-diagnostic scans, particularly in patients with arrhythmias. METHODS: A turboFLASH saturation recovery pulse sequence was used without any ECG triggering. One saturation pulse followed by 4-5 slices of undersampled radial k-space images was acquired rapidly, on the order of 40-50 msec per image. The acquisition of the set of 4-5 slices was continuously repeated approximately 4 times per second. An iterative constrained reconstruction method was used to reconstruct the ungated images. The ungated perfusion images were post-processed into three different sets of images (ungated, self-gated to near systole, and self-gated to near diastole). To test the ungated approach and compare the different processing methods, 8 patients scheduled for coronary angiography underwent stress and rest perfusion imaging with the ungated acquisition. Six patients had a history of atrial fibrillation (AF). Three blinded readers assessed image quality and presence/absence of disease. RESULTS: All 8 subjects successfully completed the perfusion CMR protocol and 7/8 underwent coronary angiography. Three patients were in atrial fibrillation during CMR. Overall, the CMR images were of high quality as assessed by the three readers. There was little difference in image quality between patients in AF compared to those in sinus rhythm (3.6±0.7 vs. 3.3±0.5). Stress/rest perfusion imaging showed normal perfusion in 4 patients, fixed perfusion defects in 2 patients, and reversible perfusion defects in 2 patients, corresponding with angiographic results. Pooled results from the independent readers gave a sensitivity of 0.92 (CI 0.65-0.99) and specificity of 0.92 (CI 0.65-0.99) for the detection of coronary artery disease using ungated perfusion imaging. The same sensitivity, and a specificity of 1 (CI 0.76-1), was achieved when the images were self-gated after acquisition into near systole or near diastole. CONCLUSIONS: Ungated radial dynamic perfusion CMR can give high quality imaging in patients in sinus rhythm and during atrial fibrillation. In this small cohort, high diagnostic accuracy was possible with this rapid perfusion imaging sequence. An ungated approach simplifies the acquisition and could expand the role of perfusion CMR to include patients with arrhythmia and those with gating problems.
No takes yet. Share an insight, caveat, or question.
Adluru et al. (2013) conducted a cohort in Suspected or established coronary artery disease (including atrial fibrillation) (n=8). Rapid ungated and self-gated radial myocardial perfusion cardiovascular magnetic resonance (CMR) with adenosine stress and gadoteridol vs. X-ray coronary angiography (or late gadolinium enhancement in 1 patient) and comparison across 3 reconstruction methods (ungated, self-gated to near-systole, self-gated to near-diastole) was evaluated on Diagnostic accuracy (sensitivity and specificity) for detection of coronary artery disease using ungated perfusion imaging (Sensitivity 0.92, Specificity 0.92, 95% CI 0.65-0.99). Ungated radial dynamic perfusion CMR achieved a sensitivity of 0.92 (95% CI 0.65-0.99) and specificity of 0.92 (95% CI 0.65-0.99) for detecting coronary artery disease.
Synapse has enriched 10 related papers. Consider them for comparative context: