Key result
Cardiac-gated intracranial elastance measurement markedly reduces measurement variance compared with non-gated approaches in swine.
Why the study?
Methodological difficulties have limited widespread clinical use of intracranial elastance measurement despite its importance in assessing risk of intracranial pressure elevation.
Answers are grounded in this paper and cite their sources.
New elastance method may flag ICP vulnerability; leaves open prospective validation before clinical use.
OBJECTIVE: Previous studies have demonstrated the importance of intracranial elastance; however, methodological difficulties have limited widespread clinical use. Measuring elastance may offer potential benefit in helping to identify patients at risk for untoward intracranial pressure (ICP) elevation from small rises in intracranial volume. The authors sought to develop an easily used method that accounts for the changing ICP that occurs over a cardiac cycle and to assess this method in a large-animal model over a broad range of ICPs. METHODS: The authors used their previously described cardiac-gated intracranial balloon pump and swine model of cerebral edema. In the present experiment they measured elastance at 4 points along the cardiac cycle-early systole, peak systole, mid-diastole, and end diastole-by using rapid balloon inflation to 1 ml over an ICP range of 10-30 mm Hg. RESULTS: The authors studied 7 swine with increasing cerebral edema. Intracranial elastance rose progressively with increasing ICP. Peak-systolic and end-diastolic elastance demonstrated the most consistent rise in elastance as ICP increased. Cardiac-gated elastance measurements had markedly lower variance within swine compared with non-cardiac-gated measures. The slope of the ICP-elastance curve differed between swine. At ICP between 20 and 25 mm Hg, elastance varied between 8.7 and 15.8 mm Hg/ml, indicating that ICP alone cannot accurately predict intracranial elastance. CONCLUSIONS: Measuring intracranial elastance in a cardiac-gated manner is feasible and may offer an improved precision of measure. The authors' preliminary data suggest that because elastance values may vary at similar ICP levels, ICP alone may not necessarily best reflect the state of intracranial volume reserve capacity. Paired ICP-elastance measurements may offer benefit as an adjunct "early warning monitor" alerting to the risk of untoward ICP elevation in brain-injured patients that is induced by small increases in intracranial volume.
No takes yet. Share an insight, caveat, or question.
Doron et al. (2020) studied Cerebral edema and raised intracranial pressure (n=7). Cardiac-gated intracranial elastance measurement vs. Non-cardiac-gated measures was evaluated on Intracranial elastance variance. Cardiac-gated intracranial elastance measurements demonstrated markedly lower variance compared with non-cardiac-gated measures in a swine model of cerebral edema.
Synapse has enriched 10 related papers. Consider them for comparative context: