Key result
Hemodialysis machine pressure signals reliably detect premature ventricular beats with ~0.9 accuracy when pulse pressure exceeds 1 mmHg.
Why the study?
Monitoring ventricular premature beats in hemodialysis patients can provide information on cardiovascular instability and electrolyte imbalance, but detection methods based on hemodialysis machine pressure signals are not established.
Observational (n=7)
No
Answers are grounded in this paper and cite their sources.
May enable VPB monitoring during hemodialysis without added sensors; extends signal methods but leaves open validation needs.
Monitoring of ventricular premature beats (VPBs), being abundant in hemodialysis patients, can provide information on cardiovascular instability and electrolyte imbalance. In this paper, we describe a method for VPB detection which explores the signals acquired from the arterial and the venous pressure sensors, located in the extracorporeal blood circuit of a hemodialysis machine. The pressure signals are mainly composed of a pump component and a cardiac component. The cardiac component, severely overshadowed by the pump component, is estimated from the pressure signals using an earlier described iterative method. A set of simple features is extracted, and linear discriminant analysis is performed to classify beats as either normal or ventricular premature. Performance is evaluated on signals from nine hemodialysis treatments, using leave-one-out crossvalidation. The simultaneously recorded and annotated photoplethysmographic signal serves as the reference signal, with a total of 149,686 normal beats and 3574 VPBs. The results show that VPBs can be reliably detected, quantified by a Youden's J statistic of 0.9, for average cardiac pulse pressures exceeding 1 mmHg; for lower pressures, the J statistic drops to 0.55. It is concluded that the cardiac pressure signal is suitable for VPB detection, provided that the average cardiac pulse pressure exceeds 1 mmHg.
No takes yet. Share an insight, caveat, or question.
Holmer et al. (2017) conducted an observational in Kidney failure undergoing hemodialysis (n=7). VPB detection based on cardiac pressure signal vs. Photoplethysmographic (PPG) signal was evaluated on Detection of ventricular premature beats (Youden's J statistic) (Youden's J statistic 0.9). A novel method using cardiac pressure signals from a hemodialysis machine reliably detected ventricular premature beats, achieving a Youden's J statistic of 0.9 when average cardiac pulse pressure exceeded 1 mmHg.