Key result
Laser sheath extraction of chronic active-fixation coronary sinus lead causes severe vascular degloving injury.
Why the study?
Transcutaneous lead extraction can cause significant morbidity and mortality, and the risk of concomitant arterial and venous injury during extraction is rare and not well documented.
Case Report (n=1)
No
Urges caution with laser sheath extraction of chronic CS leads; leaves open safer removal strategies.
Transcutaneous lead extraction can be associated with significant morbidity and mortality. The risk of causing concomitant arterial and venous injury is rare. We report a case of marginal artery rupture with coronary sinus rupture after a CS lead extraction. A 71-year-old male was admitted for extraction of a 6-year-old implantable cardioverter-defibrillator lead due to fracture from insulation break. During the lead extraction, blood pressure fell precipitously and echocardiographic findings were consistent with pericardial effusion. After unsuccessful pericardiocentesis, open chest sternotomy and evacuation of hematoma was performed. Subsequent surgical repair of several injuries was completed including the distal coronary sinus, a large degloving injury of posterior portion of the heart, and first obtuse marginal branch bleed. This case demonstrates that when performing transcutaneous lead extraction (TLE) with laser sheath, a degloving injury can cause arterial rupture with concomitant coronary sinus injury. A multidisciplinary team-based approach can ensure patient safety. Learning Objective. Implantable cardioverter-defibrillator leads will falter over time. With the advancement of new technology for extraction more frequent and serious complications will occur. Active fixation CS leads present unique challenges. In the presence of hemodynamic changes during extraction the occurrence of both an arterial and venous injury must be considered.
No takes yet. Share an insight, caveat, or question.
Green et al. (2016) conducted a case report in Fractured implantable cardioverter-defibrillator (ICD) lead (n=1). Transcutaneous lead extraction with laser sheath was evaluated. Transcutaneous extraction of a chronically implanted active fixation coronary sinus lead using a laser sheath caused a severe degloving injury with arterial and venous ruptures.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: