This is the authors' abstract. We don't add key points for this paper.
An ultraviolet Thomson-scattering system has been designed and implemented on the Shenguang-III laser facility, a 48-beam, 3ω (351 nm), 180 kJ-level laser driver for high energy density physics and inertial confinement fusion researches. The 4ω (263.3 nm) probe beam of the Thomson-scattering system is injected from the north pole (top) of the target chamber, with an assistant beam-pointing monitor to achieve high pointing accuracy. The Thomson-scattered light is collected by a double-Cassegrain optical transmission system, which provides an achromatic image over a wide wavelength range of 200–800 nm. A novel on-line alignment method is developed and applied to the diagnostic system, ensuring a volumetric positioning accuracy of ∼30 μm for the scattering volume. An online calibration is also conducted to provide the wavelength benchmark and the spectral resolution of the system. This Thomson-scattering system has been tested in a complicated experimental environment with gas-filled hohlraums, and a high-quality ion feature of the scattered light has been obtained.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Zhao et al. (2018) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: