A new method based on the coprecipitation procedure (CP) for separation/speciation using fl ame atomic absorption spectrometric (FAAS) determination was proposed for the determination of Cr(III)/ Cr(VI) in some real samples with Co(II)/α-Benzoin oxime precipitate.It enabled a very selective, basic and rapid method for chromium determination to be developed.The analytical variables including pH, amount of α-Benzoin oxime, amount of Co(II) as carrier element, and sample volume were investigated for the quantitative recoveries Cr(III)/Cr(VI).No interfering effects were observed from the concomitant ions when present in real samples.Cr(III) was quantitatively recovered on Co(II)/α-Benzoin oxime precipitate at a pH range of 7-10, while Cr(VI) was not quantitatively recovered for any of the pHs.The calibration graph was linear in the range of 1.0-10.0mg L -1 .Under the optimized conditions, detection limits of 0.01 μg L -1 and the relative standard deviations of 1.1% were found.The preconcentration factor was found to be 1000.The precipitate was dissolved in 0.1 mL of concentrated HNO 3 , and made up to 1.0 mL with deionized ultra pure water.The validation of the presented CP was checked by the analysis of certifi ed reference materials.The method was applied for the determination of Cr(III)/Cr(VI) in real samples such as natural waters and some dairy food samples, and good results were obtained (relative standard deviations <2%, recoveries >95%).