Answers are grounded in this paper and cite their sources.
Reflective objectives (ROs) mitigate chromatic aberration across a broad wavelength range. Yet, a systematic performance characterisation of ROs has not been done. In this paper, we compare the performance of a 0.5 numerical-aperture (NA) reflective objective (RO) with a 0.55 NA standard glass objective (SO), using two-photon fluorescence (TPF) and second-harmonic generation (SHG). For experiments spanning ∼1 octave in the visible and NIR wavelengths, the SO leads to defocusing errors of 25-40% for TPF images of subdiffraction fluorescent beads and 10-12% for SHG images of collagen fibres. The corresponding error for the RO is ∼4% for both imaging modalities. This work emphasises the potential utility of ROs for multimodal multiphoton microscopy applications.
No takes yet. Share an insight, caveat, or question.
Kabir et al. (2018) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: