This is the authors' abstract. We don't add key points for this paper.
PURPOSE: Starting from the implementation of a single-slice FEXI sequence, a 3D method was developed to accelerate a FEXI experiment and acquire a multi-slice volume. METHODS: 3D multi-slab FEXI divides the volume of interest into different slabs. In each slab, the slab-selective diffusion and exchange modules are combined with slice-selective detection modules. The 3D multi-slab FEXI acquisition speed was assessed against single- and multi-slice FEXI. In addition, the new method was compared to single-slice FEXI on a yeast phantom. RESULTS: 3D multi-slab FEXI acquires thirty slices within the same scan time required for single- and multi-slice FEXI to acquire one and five slices, respectively. 3D multi-slab FEXI found AXR values, homogeneity, repeatability, and reproducibility comparable to those of single-slice FEXI, increasing SNR efficiency. CONCLUSION: 3D multi-slab FEXI accelerates a FEXI experiment and acquires a volume of several slices in the same scan time as the single-slice FEXI protocol, giving comparable AXRs.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Frignani et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: