Answers are grounded in this paper and cite their sources.
We describe a simplified isocratic HPLC method for the simultaneous determination of all-trans-retinol, 3 tocopherols (α-, δ-, and γ- tocopherol), and 8 carotenoids (lutein, zeaxanthin, canthaxanthin, β-cryptoxanthin, all-trans-lycopene, 5-cis-lycopne, α-carotene, and β-carotene) in human plasma, which uses a single C18 reversed-phase column (4.6 × 250 mm, 3 µm). A photodiode array detector was used to measure the UV-vis wavelength absorbance of retinol and the eight carotenoids, and fluorescence detection was used for the tocopherols. The linear ranges of the calibration curves for the calibration solutions injected into the HPLC column are 0.02–6.0 µg/mL for all-trans-retinol, β-cryptoxanthin and α-carotene; 0.01–3.0 µg/mL for δ-tocopherol, lutein, and lycopene; 0.08–24.0 µg/mL for γ-tocopherol; 0.3–90.0 µg/mL for α-tocopherol; 0.005-1.5 µg/mL for zeaxanthin and canthaxanthin; and 0.04–12.0 µg/mL for β-carotene.