Key result
Large-lattice pulsed-field ablation yields ~82% 1-year freedom from atrial arrhythmias.
Why the study?
Most single-shot pulsed-field ablation catheters require extensive repositioning for pulmonary vein isolation, challenging contiguous, durable lesion creation.
Does a large-lattice single-shot pulsed-field ablation catheter safely and effectively achieve 1-year freedom from atrial arrhythmias in patients with paroxysmal atrial fibrillation?
Does a large-lattice single-shot pulsed-field ablation catheter safely and effectively achieve 1-year freedom from atrial arrhythmias in patients with paroxysmal atrial fibrillation?
The conformable, single-shot Sphere-360 pulsed-field ablation catheter is safe and effective for treating paroxysmal atrial fibrillation, achieving high rates of acute isolation, procedural efficiency, and 1-year clinical success.
Supports single-shot PFA efficiency and durability in PAF; leaves open confirmation via randomized trials before practice change.
BACKGROUND: Most single-shot pulsed-field ablation (PFA) catheters require extensive repositioning for pulmonary vein isolation (PVI), posing a challenge for obtaining contiguous, durable lesions. OBJECTIVE: To determine 1-year outcomes of a single-shot, all-in-one mapping and ablation PFA catheter for treating paroxysmal atrial fibrillation (PAF). METHODS: After PVI with the large-lattice catheter with expandable tip (Sphere-360), follow-up included Holter monitoring at 180 and 365 days and scheduled/symptomatic trans-telephonic monitoring (TTM) or modeled insertable loop recorder (ILR) data. Efficacy outcomes were acute PVI and 12-month freedom from atrial arrhythmias (AA), after 90-day blanking. Optional invasive remapping at 75 days facilitated waveform refinement from PULSE1, PULSE2, to the optimized PULSE3. RESULTS: At 3 centers, 100 PAF patients underwent PFA with PULSE1 (n = 30), PULSE2 (n = 20), or PULSE3 (n = 50). Procedure, left atrial dwell, and fluoroscopy times were 57.9 ± 20.6, 22.2 ± 11.8 and 6.8 ± 5.7 minutes, respectively. All 395 targeted PVs were acutely isolated, with a transpired PVI time of 11.5 ± 6.0 minutes, using 4.0 ± 1.3 lesions/PV. There were no primary safety events (serious device-related events within 7 days post-PFA). PVI durability with PULSE3 (n = 40) was 98% (per-vein) and 93% (per-patient). One-year freedom from AA recurrence was 82.0% (95% CI:73.0%-88.3%) overall, and 88.0% (95%CI, 75.2%-94.4%) for PULSE3 patients. Of the ILR sub-cohort (n = 15 PULSE3 patients), 3 patients (20%) had recurrences, with an AA burden reduction from 26% (baseline) to 1.6% (post-ablation). CONCLUSION: The large lattice PFA catheter was efficient, safe, and effective in treating PAF. The observed high PVI durability translated to clinical effectiveness, even in continuously monitored patients.
No takes yet. Share an insight, caveat, or question.
Reddy et al. (2025) studied Paroxysmal atrial fibrillation (PAF) (n=100). Single-shot, all-in-one mapping and ablation PFA catheter (Sphere-360) was evaluated on 12-month freedom from atrial arrhythmias (AA) after 90-day blanking (95% CI 73.0-88.3). Treatment with a large-lattice pulsed-field ablation catheter yielded 82.0% (95% CI: 73.0-88.3%) 1-year freedom from atrial arrhythmias overall and 88.0% for the optimized PULSE3 waveform.