A potentiometric flow injection system was adopted by using a Li+‐selective electrode with improved selectivity for sensitive and selective determination of Li+ in human serum spiked samples. For the optimization of the offered flow injection system, multivariate method was utilized. Potentiometric characteristics of the system were investigated (linear range: 5.0×10−5−1.0×10−1 mol L−1, slope: 56.93 mV per decade, detection limit: 3.4×10−5 mol L−1, sampling‐rate: 40 injections per hour). The detection of Li+ in real serum samples was achieved. The obtained results for the spiked serum samples were in good agreement with IC results at 95 % confidence level.