Key result
Updated continuous ECG monitor linked to ~14% higher arrhythmia yield versus its predecessor.
Why the study?
The impact of device-specific differences in wear performance and analyzable ECG duration on arrhythmia yield has not been well characterized at a population level.
Does an updated 14-day patch-based continuous monitor improve wear duration and arrhythmia yield compared to its predecessor in patients requiring ambulatory ECG monitoring?
Cohort (n=911,819)
Does an updated 14-day patch-based continuous monitor improve wear duration and arrhythmia yield compared to its predecessor in patients requiring ambulatory ECG monitoring?
Odds Ratio: 1.14 (95% CI 1.12–1.15)
Absolute Event Rate: 78.5% vs 76.9%
An updated 14-day patch-based continuous monitor reduced early discontinuations, extended analyzable ECG duration, and improved arrhythmia detection yield at a population level compared to its predecessor.
Longer wear times were associated with higher arrhythmia yield in retrospective data; leaves open whether next-generation patches should supplant predecessors in practice.
Background The impact of device-specific differences in wear performance and analyzable ECG duration on arrhythmia yield has not been well characterized at a population level. We evaluated device performance and arrhythmia yield in a large post-market comparison of two long-term continuous monitors. Methods We performed a retrospective study comparing the post-market performance of an updated patch-based long-term continuous monitor (LTCM; iRhythm Technologies, San Francisco, CA) with its predecessor among patients prescribed 14-day monitoring. Both devices used the same FDA-cleared algorithm and technician review workflow. Outcomes included wear performance, analyzable ECG duration, arrhythmia yield (atrial fibrillation, supraventricular tachycardia, ventricular tachycardia, pauses, and atrioventricular block), skin irritation, and patient satisfaction. Results In 911,819 patients (age 61.8 ± 17.7 years, 53.7% female), the new LTCM (n=446,344) had longer wear time (mean 12.8 ± 2.6 vs. 12.1 ± 3.3 days, p<0.001) and fewer early discontinuations (1.0% vs. 2.3%, p<0.001) than its predecessor (n=465,475). After adjustment for age and sex, the new LTCM was associated with higher arrhythmia yield (78.5% vs. 76.9%; OR 1.14, 95% CI 1.12-1.15), with higher detection of actionable arrhythmias, AF, SVT, and VT. Across both devices, every additional 0.5 days of analyzable ECG duration was associated with a 6.6% increase in adjusted odds of arrhythmia detection (OR 1.066, 95% CI 1.065-1.067). Conclusions A new LTCM demonstrated fewer early wear discontinuations, longer wear duration, and higher arrhythmia yield than its predecessor. Modest increases in analyzable ECG duration were associated with higher arrhythmia yield, suggesting that small improvements in wear performance can have measurable population-level effects.
No takes yet. Share an insight, caveat, or question.
Knight et al. (2026) conducted a cohort in Indication for ambulatory ECG monitoring (n=911,819). Updated patch-based long-term continuous monitor (LTCM) vs. Predecessor LTCM was evaluated on Arrhythmia yield (OR 1.14, 95% CI 1.12-1.15). An updated long-term continuous monitor was associated with a higher arrhythmia yield compared to its predecessor (78.5% vs. 76.9%; OR 1.14; 95% CI 1.12-1.15).