Key result
Insertable loop recorders show undersensing in ~36% of auto-recorded episodes, unaffected by body posture.
Why the study?
The newer insertable loop recorder device with auto-activation capabilities may encounter sensitivity problems such as oversensing or undersensing, and the influence of body position on sensing is unclear.
Observational (n=32)
AIMS: The newer insertable loop recorder device (Reveal Plus 9526, Medtronic Inc, Minneapolis MN, U.S.A.) is equipped with auto-activation capabilities. In this study we investigated whether the new device encounters sensitivity problems in the form of oversensing or undersensing. We also tested whether body position changes influence the sensed electrogram's amplitude, therefore affecting the device's autosensing ability. METHODS AND RESULTS: We enrolled 32 patients aged 58.1+/-11.7 years with a loop recorder implanted either in the left parasternal or in the heart's apex area, to investigate undiagnosed syncopal episodes after initial assessment. During the follow-up period (5.7+/-2.7 months), that began 6 months after the device implantation, we analysed 284 auto-recorded episodes (103 undersensed and 181 correctly sensed). No oversensing episodes were recorded. THE RECORDED QRS amplitude in five different body positions: supine, left and right lateral, sitting and standing was measured. No statistically significant changes were observed in the different postures. Even when patients were divided into subgroups according to implant site or the occurrence of undersensing episodes, no significant inter- or intra-group differences of the sensed electrogram were observed. CONCLUSION: Undoubtedly this device is the current gold standard for the diagnosis of unexplained syncope. In the population we studied though, the device faced undersensing problems. Consequently, the device should be equipped with more advanced software for QRS morphology recognition, to improve the diagnostic accuracy of the recorder.
No takes yet. Share an insight, caveat, or question.
Chrysostomakis et al. (2003) conducted an observational in Undiagnosed syncopal episodes (n=32). Insertable loop recorder device (Reveal Plus 9526) was evaluated on Sensitivity problems (oversensing or undersensing) and sensed electrogram QRS amplitude changes across body positions. The insertable loop recorder device faced undersensing problems, with 103 of 284 auto-recorded episodes (36.3%) undersensed, though body posture did not significantly affect QRS amplitude.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: