Key result
HAART aortic annuloplasty ring achieves 100% survival and valve competence in preclinical calf models.
Why the study?
A three-dimensional aortic annuloplasty ring was developed for intra-annular placement to increase applicability and stability of aortic valve repair, requiring in vivo safety testing.
Answers are grounded in this paper and cite their sources.
Device safe in calf model; leaves open human translation for aortic valve repair.
OBJECTIVE: To increase applicability and stability of aortic valve repair, a three-dimensional aortic annuloplasty ring has been developed for intra-annular placement. The goal of this study was to test the safety of this device with in vivo implantation in the calf model. METHODS: In 10 chronic calves, the HAART annuloplasty ring was sutured to the aortic valve annulus using cardiopulmonary bypass. The animals were recovered and followed for 1-2 months. Serial echocardiography was used to evaluate valve competence, and contrast aortograms and CT angiograms were obtained in selected animals. After completion of follow-up, each animal was euthanized, and aortic endoscopy was performed under water distension in five. Full autopsies with histologic examinations were performed. RESULTS: All animals survived surgery. Two were euthanized in the first week for complications, and the remaining eight calves were followed uneventfully for the 1-2 months. Serial echocardiography showed completely competent valves in all but one animal, in which the ring was intentionally up-sized to test the sizing strategy. Contrast aortographic and CT angiographic findings were similar to the echocardiograms. Postmortem examination showed proper seating of all rings with endothelialization at 1-2 months. All valves demonstrated good leaflet coaptation and no abnormalities. CONCLUSIONS: In vivo testing of a three-dimensional aortic annuloplasty ring in a chronic calf model proved to be very successful and safe. Using the sizing and implant strategies developed, human trials seem indicated.
No takes yet. Share an insight, caveat, or question.
Rankin et al. (2012) studied Aortic valve repair evaluation (n=10). HAART three-dimensional aortic annuloplasty ring was evaluated on Safety and valve competence. Implantation of the HAART three-dimensional aortic annuloplasty ring in 10 calves resulted in 100% surgical survival and competent valves in all appropriately sized rings after 1 to 2 months.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: