In this paper, a fluorescent aptamer sensor based on magnetic ferroferric oxide@boron, nitrogen‐doped graphene quantum dots (Fe 3 O 4 @B,N‐GQDs) was developed to detect Golgi protein 73 (GP73). GP73 aptamer (GP73 Apt ) functioned with Fe 3 O 4 @B,N‐GQDs (Fe 3 O 4 @B,N‐GQDs‐GP73 Apt ) and was used as the fluorescent donor, while molybdenum disulfide (MoS 2 ) with a large surface served as the fluorescent acceptor. The fluorescence of Fe 3 O 4 @B,N‐GQDs‐GP73 Apt was quenched by MoS 2 based on the fluorescence resonance energy transfer (FRET) principle. However, in the presence of GP73, Fe 3 O 4 @B,N‐GQDs‐GP73 Apt could specifically bind to GP73, forming a Fe 3 O 4 @B,N‐GQDs‐GP73 Apt ‐GP73 complex. This binding event caused Fe 3 O 4 @B,N‐GQDs‐GP73 Apt to move away from the surface of MoS 2 , thus blocking the FRET process and recovering the fluorescence. Under optimal conditions, a linear relationship was established between the fluorescence recovery and the concentration of GP73 within the range of 10–1000 ng/mL ( R 2 = 0.9918), and the limit of detection was 7.37 ng/mL. Additionally, when the sensor was applied to test actual samples, the recovery rates were in the range of 98.80%–101.18%, and the relative standard deviations were between 0.29%–3.04%. These findings demonstrated the excellent detection performance of the proposed fluorescent aptamer sensor.