Key result
GE Marquette 12SL ECG algorithm shows high overall cardiologist agreement but varies by age and sex.
Why the study?
Automated 12-lead ECG algorithms are widely used but their disagreement with cardiologists across patient subgroups is rarely audited on public data.
Does the GE Marquette 12SL automated ECG algorithm agree with expert cardiologists in diagnosing MI, conduction disturbances, and LVH across different age and sex subgroups?
Observational (n=13,480)
Does the GE Marquette 12SL automated ECG algorithm agree with expert cardiologists in diagnosing MI, conduction disturbances, and LVH across different age and sex subgroups?
A widely deployed commercial ECG interpretation algorithm shows significant age- and sex-dependent biases in diagnosis compared to expert cardiologists, highlighting the necessity for continuous, demographically stratified post-deployment auditing.
Answers are grounded in this paper and cite their sources.
Automated CD under-detection by 12SL versus experts warrants manual review in practice; leaves open needs for age/sex-specific algorithm refinement and prospective validation.
BACKGROUND: Automated 12-lead ECG algorithms are widely used, yet their disagreement with cardiologists across patient subgroups is rarely audited on public data. We characterized the directional discordance of the GE Marquette 12SL algorithm against expert labels. Methods: In this retrospective audit of 14,885 ECGs from 13,480 patients (PTB-XL/PTB-XL+ dataset), we compared 12SL (v23) statements with standardized cardiologist labels for myocardial infarction (MI), conduction disturbance (CD), and left ventricular hypertrophy (LVH). We calculated observed agreement and paired marginal differences with 95% confidence intervals. Discordance was modeled using generalized estimating equations incorporating age, sex, and their interaction. Results: Observed agreement was high (0.842-0.896) but directionally divergent. 12SL produced fewer positive statements than cardiologists for CD (paired marginal difference 0.092, 95% CI 0.085-0.098) and more for LVH (-0.070, 95% CI -0.075 to -0.064); MI showed no net difference. The false-negative proportion for CD decreased with age, while false-positive proportions rose with age for CD and MI. For LVH false positives, the age association was strongly modified by sex (interaction 1.60, 95% CI 1.47-1.73), being absent in women but steep in men. The LVH direction was specific to primary 12SL statements and reversed if supportive statements were included. Conclusion: A widely deployed commercial ECG algorithm exhibits diagnosis-specific, directional discordance with expert reading that varies by age and sex, underscoring the value of reproducible post-deployment auditing. Because no clinical gold standard was available, these findings describe agreement with experts rather than absolute diagnostic accuracy.
No takes yet. Share an insight, caveat, or question.
Du et al. (2026) conducted an observational in Electrocardiogram interpretation (Myocardial Infarction, Conduction Disturbance, Left Ventricular Hypertrophy) (n=13,480). GE Marquette 12SL algorithm (v23) vs. Expert cardiologists was evaluated on Paired marginal difference for conduction disturbance (CD) (paired marginal difference 0.092, 95% CI 0.085-0.098). The GE Marquette 12SL algorithm demonstrated high overall agreement with cardiologists but exhibited diagnosis-specific directional discordance that varied by age and sex.
Synapse has enriched 10 related papers. Consider them for comparative context: